MRI Safe for Patients with Implanted Legacy Devices

By HospiMedica International staff writers
Posted on 09 Jan 2018
Patients with a legacy pacemaker or implantable cardioverter-defibrillator (ICD) system that need to undergo a magnetic resonance imaging (MRI) scan have nothing to fear, according to a new study.

Researchers at the University of Pennsylvania (Penn; Philadelphia, USA) and Johns Hopkins University (JHU; Baltimore, MD, USA) conducted a prospective, nonrandomized study to assess the safety of MRI at a magnetic field strength of 1.5 Tesla in 1,509 patients who had a pacemaker (58%) or an ICD (42%) that was not considered to be MRI-conditional (also known as a legacy device). More...
Overall, the patients underwent 2,103 thoracic and non-thoracic MRI examinations that were deemed to be clinically necessary.

The pacing mode was changed to asynchronous mode for pacing-dependent patients, and to demand mode for other patients, with tachyarrhythmia functions disabled. Outcome assessments included adverse events and changes in device parameters indicating lead and generator function and interaction with surrounding tissue. The results showed no reports of long-term clinically significant adverse events. Following the MRI scan, the patient's device reset to a backup mode in nine cases (0.4%), of which eight were transient. One pacemaker reset to ventricular inhibited pacing and could not be reprogrammed.

A decrease in P-wave amplitude was the most common notable change in device parameters (occurring in one percent of patients) immediately after MRI. At long-term follow-up, the most common notable changes from baseline were decreases in P-wave amplitude (4%), increases in atrial capture threshold (4%), increases in right ventricular capture threshold (4%), and increases in left ventricular capture threshold (3%). The study was published on December 28, 2017, in the New England Journal of Medicine (NEJM).

“Patients who have pacemakers or defibrillators are often denied the opportunity to undergo magnetic resonance imaging because of safety concerns, unless the devices meet certain criteria specified by the FDA, termed MRI-conditional devices,” concluded lead author Saman Nazarian, MD, PhD, of the University of Pennsylvania, and colleagues. “No long-term clinically significant adverse events were reported, and the observed changes in lead parameters were not clinically significant, and did not require device revision or reprogramming.”

Until now, patients with ICD systems have been contraindicated from receiving MRI scans because of potential interactions between the MRI and device function, which might result in risk to patients. The restriction has resulted in a critical unmet need as data have shown that within four years, 36% percent patients with ICDs in the United States are likely to need an MRI.

Related Links:
University of Pennsylvania
Johns Hopkins University

Platinum Member
Real-Time Diagnostics Onscreen Viewer
GEMweb Live
Gold Member
Enteral Feeding Pump
SENTINELplus
Morcellator
TCM 3000 BL
Premium Air-Mattress
MA-51
Read the full article by registering today, it's FREE! It's Free!
Register now for FREE to HospiMedica.com and get access to news and events that shape the world of Hospital Medicine.
  • Free digital version edition of HospiMedica International sent by email on regular basis
  • Free print version of HospiMedica International magazine (available only outside USA and Canada).
  • Free and unlimited access to back issues of HospiMedica International in digital format
  • Free HospiMedica International Newsletter sent every week containing the latest news
  • Free breaking news sent via email
  • Free access to Events Calendar
  • Free access to LinkXpress new product services
  • REGISTRATION IS FREE AND EASY!
Click here to Register








Channels

Surgical Techniques

view channel
Image: Professor Bumsoo Han and postdoctoral researcher Sae Rome Choi of Illinois co-authored a study on using DNA origami to enhance imaging of dense pancreatic tissue (Photo courtesy of Fred Zwicky/University of Illinois Urbana-Champaign)

DNA Origami Improves Imaging of Dense Pancreatic Tissue for Cancer Detection and Treatment

One of the challenges of fighting pancreatic cancer is finding ways to penetrate the organ’s dense tissue to define the margins between malignant and normal tissue. Now, a new study uses DNA origami structures... Read more

Patient Care

view channel
Image: The portable biosensor platform uses printed electrochemical sensors for the rapid, selective detection of Staphylococcus aureus (Photo courtesy of AIMPLAS)

Portable Biosensor Platform to Reduce Hospital-Acquired Infections

Approximately 4 million patients in the European Union acquire healthcare-associated infections (HAIs) or nosocomial infections each year, with around 37,000 deaths directly resulting from these infections,... Read more
Copyright © 2000-2026 Globetech Media. All rights reserved.