Features Partner Sites Information LinkXpress hp
Sign In
Advertise with Us
Radcal IBA  Group

Download Mobile App




Number of Stent Connectors Could Explain Longitudinal Weakness

By HospiMedica International staff writers
Posted on 13 Dec 2011
It may be the number of connectors in a stent that determines whether or not it is more susceptible to longitudinal compression, according to a new study.

Researchers at Mercy Ascot Hospital (Auckland, New Zealand) conducted a bench analysis of seven different stents, which involved the compression of the stents by 5 mm. More...
The researchers found that the least amount of force was required to compress the Omega (Boston Scientific) and Driver (Medtronic) stents, each requiring 0.40 N and 0.71 N of force, respectively. The Integrity (Medtronic), Liberté (Boston Scientific), Vision (Abbott Vascular), and MultiLink (Abbott Vascular) stents all required significantly more force to compress 5 mm, when compared with the Omega stent. The soon-to-be-discontinued Cypher stent (Cordis; Bridgewater Township, NJ, USA) was the hardest to compress, requiring 1.33 N. Conversely, when 0.5 N of force was applied to the Cypher stent it did not shorten, while little shortening was observed with the Vision and MultiLink stents, and the most distortion with the Driver and Omega stents.

When the researchers applied force to elongate the stents by 1 mm, the Omega and Driver stents were more easily deformed, while the force needed to elongate the Liberté and Integrity stents was similar and significantly greater than the force needed to alter the structure of the Driver stent. All the stents were equivalent in terms of the amount of force needed to stretch the stent a full 5 mm, with exception of the Cypher stent, which required significantly more force. The researchers suggest that a stent design change ensuring three connectors, especially at the proximal end of a stent, should increase longitudinal integrity, but perhaps at the expense of stent flexibility. The study was published in the December 2011 issue of Cardiovascular Interventions.

“Stent deliverability, strongly influenced by flexibility, is the property that cardiologists desire most, and designs that have high longitudinal integrity may not have high flexibility and deliverability,” concluded lead author John Ormiston, MBChB, and colleagues. “We show that the earlier-generation Cypher Select, a six-connector design, had the greatest resistance to longitudinal distorting forces. This stent is no longer widely used because other designs have better flexibility, deliverability, crossing profile, radiopacity, side-branch access, and freedom from strut fracture.”
'
Stent compression is a potentially devastating phenomenon that can result in major adverse cardiac events. Some of the newer stents, in search of improved flexibility, might have compromised their longitudinal strength, compressing along their longitudinal axis, thus causing folding, wrinkling, bunching together of rings, or other forms of deformation. The use of the postdilating balloon, intravenous ultrasound (IVUS) catheter, guide catheters, or thrombectomy catheters could cause longitudinal compression.

Related Links:
Mercy Ascot Hospital
Cordis



Platinum Member
STI Test
Vivalytic Sexually Transmitted Infection (STI) Array
Gold Member
Enteral Feeding Pump
SENTINELplus
Digital X-Ray Detector Panel
Acuity G4
Newborn Hearing Screener
ALGO 7i
Read the full article by registering today, it's FREE! It's Free!
Register now for FREE to HospiMedica.com and get access to news and events that shape the world of Hospital Medicine.
  • Free digital version edition of HospiMedica International sent by email on regular basis
  • Free print version of HospiMedica International magazine (available only outside USA and Canada).
  • Free and unlimited access to back issues of HospiMedica International in digital format
  • Free HospiMedica International Newsletter sent every week containing the latest news
  • Free breaking news sent via email
  • Free access to Events Calendar
  • Free access to LinkXpress new product services
  • REGISTRATION IS FREE AND EASY!
Click here to Register








Channels

Surgical Techniques

view channel
Image: Professor Bumsoo Han and postdoctoral researcher Sae Rome Choi of Illinois co-authored a study on using DNA origami to enhance imaging of dense pancreatic tissue (Photo courtesy of Fred Zwicky/University of Illinois Urbana-Champaign)

DNA Origami Improves Imaging of Dense Pancreatic Tissue for Cancer Detection and Treatment

One of the challenges of fighting pancreatic cancer is finding ways to penetrate the organ’s dense tissue to define the margins between malignant and normal tissue. Now, a new study uses DNA origami structures... Read more

Patient Care

view channel
Image: The portable biosensor platform uses printed electrochemical sensors for the rapid, selective detection of Staphylococcus aureus (Photo courtesy of AIMPLAS)

Portable Biosensor Platform to Reduce Hospital-Acquired Infections

Approximately 4 million patients in the European Union acquire healthcare-associated infections (HAIs) or nosocomial infections each year, with around 37,000 deaths directly resulting from these infections,... Read more
Copyright © 2000-2026 Globetech Media. All rights reserved.